NEW! SimCube® Patient Simulation System ADVANCED – LubdubBazaar
Help Line: +91 755 000 9944
Cart 0
NEW! SimCube® Patient Simulation System ADVANCED

Pronk Technologies

NEW! SimCube® Patient Simulation System ADVANCED

Rs. 0.00

NEW! SimCube® Patient Simulation System ADVANCED

Includes: SC-5 SimCube® + OxSim Flex™ + SimSlim®

Introducing the latest patient simulation system now with OxSim Flex. This simulation system enables simultaneous testing of all vital signs parameters including ECG, Respiration, Arrhythmia, 4 Invasive Blood Pressures, NIBP, user-selectable SpO2 simulations, Cardiac Output, and YSI Temperature. Includes all features needed to service and perform preventive maintenance on any brand of vital signs monitor. Additional features include digital manometer, automated leak test mode, automatic SpO2 manufacturer detection, Masimo rainbow SET® compatibility, user-selectable SpO2, heart rate and respiration sequences for faster alarm testing. All battery operated. Save hundreds when you order as a complete system!

Thousands of Biomeds worldwide trust SimCube Vital Signs Simulation systems for portable, accurate vital signs simulations. One-button operation means ultra-fast PMs. 4-year parts and labor warranty plus durability to withstand 50 drops from 3.5 feet yield unsurpassed value to this compact package.

NIBP Simulations

  • Adult (120/80 mmHg), Neonatal (70/40 mmHg), Hypertensive (190/120 mmHg), and Hypotensive (80/40 mmHg) NIBP Simulations
  • Digital Manometer
  • Over Pressure (Peak Detect) Test mode measures the peak pressure produced by the patient monitor. The readout latches the highest pressure valve sensed, in 0.1 mmHg steps
  • Leak Rate Test Mode

ECG/Respiration Simulations

  • 12-lead ECG Simulation
  • 8 Selectable Heart Rates: 30, 60, 90, 120, 140, 160, 200, 240 bpm
  • 10 Selectable Arrhythmias: NSR w/ PVCs, VTACH, VFIB, Asys, Bigeminy, Pacer, Trigeminy, ST+, 2nd Deg. Block, Square Wave
  • 8 Selectable Respiration Rates: 15, 30, 45, 60, 70, 80, 100, 120 bpm
  • Banana Plug Interface

SpO2 Simulations

  • Saturations in 1% increments (10-100%)
  • Pulse rates in 1 BPM increments (25-240)
  • Create custom preset simulations for faster Alarm Testing
  • Perfusion Index levels in 1% increments (10-100%)
  • Factory Preset Simulations:
    • Saturation: 85%, Heart Rate: 80 bpm, Perfusion Index: approx. 2.0
    • Saturation: 95%, Heart Rate: 40 bpm, Perfusion Index: approx. 2.0
    • Saturation: 98%, Heart Rate: 80 bpm, Perfusion Index: approx. 2.0
    • Saturation: 98%, Heart Rate: 140 bpm, Perfusion Index: approx. 2.0
    • Saturation: 99%, Heart Rate: 80 bpm, Perfusion Index: approx. 2.0
  • Masimo rainbow SET® Compatible
  • Automatic detection of SpO2 Manufacturer
  • Color Display – OLED Technology

IBP Simulations

  • 4 IBP Channels
  • 8 Dynamic: 120/80, 60/30, 30/10, 20/10, 125/15, 60/5, 8/2, 4/1 mmHg
  • 7 Static: 0, 25, 50, 100, 150, 200, 250 mmHg

Cardiac Output Simulations

  • 3, 5 and 7 l/m. No adapter cable required

Temperature Simulations

  • YSI 400 and 700
  • 25, 37, and 40 degrees Celsius

Package Contents

  • SC-5 SimCube® NIBP Simulator
  • OX-2 OxSim Flex™ Optical Pulse Oximeter Simulator
  • SL-8 SimSlim® Multi-Parameter Patient Simulator
  • Battery Boost
  • (6) NIBP Adapters
  • Cuff Jacket
  • (1) IBP extension cable
  • (1) IBP Adapter Cable of your choice
  • (1) Specially-modified nylon carrying case

SC-5 SimCube® Specification:-

Physical Dimensions

Size

3″ x 3″ x 3.5″ (7.6cm X 7.6cm X 8.9cm)

Weight

2.5 Lbs

Power

External A/C Adaptor (Output: 6VDC / 2amps, 2.1mm, center positive connector) or 4 AA Batteries (with Battery Boost Option)

NIBP Connection

Quick Disconnect

ECG/Resp Connection

10 ECG snaps

  • Manometer:

Range

– 400 to + 400 mmHg

Resolution

0.1 mmHg

Accuracy

+/-1 mmHg

  • User Interface:

Operating Modes

Adult NIBP
Neo NIBP
Hypertensive NIBP
Hypotensive NIBP
Manometer
Peak Detect / Overpressure
HR Seq. Alarm Test
ECG Asystole
ECG Resp OFF
ECG Pace ON
Arrhythmia Sequence
Invasive BP Zero
Invasive BP 100, 200
Invasive BP Sequence

  • NIBP Adult Simulation:

Simulated Pressure

120/80 (97) mmHg

Simulated Heart Rate

70 bpm

Simulated Pulse Volume

1 ml

  • NIBP Neonatal Simulation:

Simulated Pressure

70/40 (51) mmHg

Simulated Heart Rate

95 bpm

Simulated Pulse Volume

0.25 ml

  • NIBP Hypertensive Simulation:

Simulated Pressure

190/120 (142) mmHg

Simulated Heart Rate

70 bpm

Simulated Pulse Volume

1 ml

  • NIBP Hypotensive Simulation:

Simulated Pressure

80/40 (58) mmHg

Simulated Heart Rate

70 bpm

Simulated Pulse Volume

1 ml

  • ECG Simulation :

Isolated

Yes

Synchronized with NIBP

Yes

R Wave Size

1mV (lead I) +/- 5%

R Wave Width

35 ms

Wave Shape

QRS wave

Connection

10 Snaps

Simulation Rates

70, 95, Asystole, Arrhythmia, Pacer, HR seq.

  • HR Sequence (Alarm Test):

Timing

30 seconds each of: 30, 60, 90, 120, 45, 160, and 220 bpm

  • Pacer Simulation:

Isolated

Yes

Synchronized with NIBP

Yes

Pacer Size

3 mV

Pacer Width

1.2 ms

  • Respiration Simulation:

Isolated

Yes

Wave Shape

Square Wave

Size

4 Ohm

Rate

20 bpm (40 bpm for neonatal)

Simulation rates

30 seconds each of: Apnea, 30, 45, 60, 22, 80, 110.

  • Arrhythmia Simulation:

Cardiac failure sequence

90 seconds of normal beats interspersed with PVCs and Runs, followed by 30 seconds of VTAC, followed by 30 seconds of VFIB, concluding with 30 seconds of asystole.

  • Peak Detect:

Precision

1mmHg

  • Invasive Blood Pressure Simulation:

Isolated

Yes

Synchronized with NIBP

Yes

Excitation Voltage

DC range = 3.3 to 5.7 AC range = 6.65 to 11.4p-p

Pressure range

0-250 mmHg

Simulated Pressure accuracy

+/- 1 mmHg

Wiring

+ Excit = pin 1, – Excit = pin 4, + Sig = pin 3, -Sig = pin 6

Simulation rates

Dynamic = 120/80, 70/40, 190/120
Static = 0, 100, 200
Step = 0,25,50,100,150,200, 250

  • Environmental:

Voltage Range

100-240 VAC, 50-60 Hz

 

OX-2 OxSim Flex™

 

Physical Dimensions

     Size

2.5″ x 2.5″ x 1.25″ (6.4 cm x 6.4 cm x 3.8 cm)
(excluding finger)

     Weight

5 oz (142 g) (excluding battery)

     SpO2 Connection

Optical

Environmental

     Operating Temperature

50° F to 95° F (10° C to 35° C).
Relative Humidity 20% to 80% non-condensing.

     Operating Altitude

0 – 9,843 ft (0 – 3,000 meters)

     Storage Temperature

-4° F to +140° F (-20° C to +60° C)

Power

     Battery Power

1 AA battery (any chemistry)

     Battery Life

10 to 12 hours using alkaline battery
60 days in sleep mode (unit sleeps after 10 min. with no probe connection)

     AC Operation

Included Class II limited power source 6VDC supply operates from 100-240VAC, 50-60Hz

Saturation

     Range

10% – 100%

     Accuracy

Saturation

Autoprobe Selected

Manufacturer Probe Selected

85% – 100%

± 2% of reading

± 1% of reading

70% – 84%

± 3% of reading

± 2% of reading

60% – 69%

NA

±3% of reading

Heart Rate

     Range

25 – 240 BPM

     Accuracy

± 2 BPM

Perfusion Index

     Range

10% – 100% (0.2 – 2.0 PI)

 

SL-8 SimSlim®

Physical Dimensions

     Size

7″x 1.25″ x 0.8″ (17cm X 3.2cm X 2.0cm)

     Weight

4 oz (0.1kg)

     ECG/Resp Connection

10 ECG snaps with 4mm and 1/8″ banana plug grooves

     IBP Connection

4 X 6 pin Mini-DIN

     YSI Temp Connection

2 X 3.5mm mini stereo jack

     C.O. Connection

Standard 3 pin catheter connection

Power

     Power Supply

3.6 VDC. 0.030mA max. Internal Lithium Thionyl Chloride battery

     Battery Life

5 to 10 year continuous usage battery life, depending on usage**

Environmental

3.6 VDC. 0.030mA max. Internal Lithium Thionyl Chloride battery

User Interface

     Single Button Operation

Double Click = Change Mode
Single Click = Show Mode/Zero IBP/Advance Waveform
Hold (2s) = Start C.O.

Operating Modes

 

MODE

ECG

HR

RR

IBP1

IBP2

IBP3

IBP4

C.O.

ADULT

NSR

60

30

120/80

30/10

125/15

8/2

OFF

NEO

NSR

140

70

60/30

20/10

60/5

4/1

3 l/m

HR STEPS

NSR

STEPS

STEPS

STEPS

STEPS

STEPS

STEPS

5 l/m

ARR MODE 1

VARIOUS

VARIOUS

OFF

100mmHg

100mmHg

100mmHg

100mmHg

7 l/m

ARR MODE 2

VARIOUS

VARIOUS

OFF

200mmHg

200mmHg

200mmHg

200mmHg

OFF

 

HR Steps

30, 60, 90, 120, 160, 200, 240 (BPM)

Resp Steps

15, 30, 45, 60, 80, 100, 120 (RPM)

IBP Steps

0, 25, 50, 100, 150, 200, 250 (mmHg)

ARR MODE 1

NSR with PVCs, VTACH, VFIB, ASYSTOLE

ARR MODE 2

BIGEMINY, PACER, TRIGEMINY, ST+, 2nd DEGREE BLOCK, SQUARE WAVE

ECG Simulation

     Isolated

Yes

     R Wave Size

1mV (lead II) +/- 10% on NSR mode only

     R Wave Width

35 ms

     Wave Shape

QRS wave

     Connection

10 Snaps, plus 4mm and 1/8” banana plug grooves

     Rates

60, 140, 30, 60, 90, 120, 160, 200, 240 (BPM)

Pacer Simulation

     Isolated

Yes

     Synchronized with ECG

Yes

     Pacer Size

0.4 to 2 mV

     Pacer Width

1.6 ms

Respiration Simulation

     Isolated

Yes

     Synchronized with NIBP

Yes

     Size

0.4 to 2 mV

     Rates

30, 70, 15, 30, 45, 60, 80, 100, 120 (RPM)

Invasive Blood Pressure Simulation

     Isolated

Yes

     Synchronized with NIBP

Yes

     Excitation Voltage

DC range = 2.5 to 5.7V AC range = 5 to 11.4Vp-p

     Pressure range

0-250 mmHg

     Simulated Pressure accuracy

+/- 0.2%

     Wiring

+ Excitation = pin 1, – Excitation = pin 4, + Sig = pin 3, -Sig = pin 6

     Simulation rates

Dynamic = 120/80, 30/10, 125/15, 8/2, 60/30, 20/10, 60/5, 4/0
Static =100, 200.
Step = 0,25,50,100,150,200, 250

Temperature Simulation

     YSI 400 Simulation Values

25, 37, 40 (degrees C)

     YSI 700 Simulation Values

25, 37, 40 (degrees C)

     Simulation Accuracy

+/- 0.03 C

CO Simulation

     Baseline Blood Temp

37.0 C

     Simulation Values

3, 5, 7 (l/m)

     Adapter Cable / Box

None Required

     Simulated Blood Temp Accuracy

+/- 0.1 C

Mechanical

     Housing Material

Carbotex Polycarbonate. 0.1″ wall thickness.

     Housing Construction

Clamshell, with 10 #4-40 fasteners all the way through both case parts and 4 plastite corner fasteners

     Attachments

Rugged key chain attachment for securing to belt, lanyard, reel, case, workbench, cart, etc.

     Internal Electrical Connections

All connections via soldered flex strip; no crimps or internal removable connectors.

     Labeling

All necessary instructions and data are printed on the case of the simulator, as well as in the operator’s manual.

 


Share this Product


More from this collection